PDJ008 Electrosurgical Unit

Category :

Description

Scope of Application

It is suitable for various surgeries on cutting or coagulation, including general surgery, heart, gynecology, anorectal surgery, orthopedics, thoracic surgery, tumor, etc. It is also used with surgeries of endoscope, hysteroscope, laparoscope, ENT endoscope, etc. The bipolar can be used in the fine surgeries of microsurgeryneurology, ENT and ophthalmology, hand surgery, etc.

 

Functional Features

● Cutting: that is pure cutting, the effect is fine cutting, tissue surface is fresh and no scab;

 

● Blend: The effect is to provide good coagulation while cutting. It is suitable for superficial coagulation of tissue in surgery and can be used with endoscopy.

 

● Coag 1: It can be called soft electrocoagulation or point coagulation. It is suitable for deeper coagulation of tissue in surgery and can be used with endoscopy. Because smaller area of action can reduce adhesion between electrode and tissue, it is better for rehabilitation of patients.

 

● Coag 2: It can be called strong electrocoagulation or surface coagulation. I t has the faster coagulation and larger area of action.

 

● Bipolar Coag: It can be called rapid bipolar electrocoagulation. It can achieve the effect of coagulation at the moment of start-up and reduce the rate of misoperation during surgery.

 

Technical Parameters:

Ambient temperature range: 10ºC~40ºC; Relative humidity range: 30%~75%;
Atmospheric pressure range: 700hpa~1060hpa; Power supply: 220V (±22V),   50Hz (±1Hz);
Working frequency: 360kHz~460   kHz; Type of equipment: CF;
The power consumption of the whole device   is less than 1000VA. (Cutting function: 400W)

Five Working Modes:

Mono cut Cutting: 400W (load 500) Mono Coag Soft Coag: 100W   (load 500)
Blend: 150W (load   500) Strong Coag: 80W   (load 500)
Bipolar Bipolar Coag: 50W   (load 200)