Flow Cytometer

Description

Item Parameters
Laser 488nm solid
Detector FSC,   SSC, FL1-FL5
 

Fluorescence   Channel

FL1:525/50(FITC)                    FL2:585/40(PE)

FL3:630/30(PE-Texas Red/ECD/PI)     FL4:697/60(PerCP/PE-Cy5/PE-Cy5.5)

FL5:780/60(PE-Cy7)

Fluorescence   sensitivity FITC   ≤ 100 MESF; PE ≤ 50 MESF
Fluorescence   linearity ≥   0.99
FSC   sensitivity ≤   1μm
Test   particle size 0.5-50μm
Carry   Over ≤   0.5%
Instrument   CV FSC   and all FL channel CV less than 2%
FSC   and SSC CV Can   separate red blood cells and platelet, can separate white blood cells into 3   groups
Test   speed ≥38000   cells/s
Minimum   Sample Vol 90μL
Measure   Range Maximum   Support 107
Threshold Double   threshold, minimum cells fraction
Fluorescence   compensation Support   auto online compensation, offline compensation and visualized compensation
Item   Template Support   Item template, acquisition condition/ fluorescence compensation 1-click use
Intake   method Manual   intake

Rotary   Wheel auto-sampling with auto- mixing function: 32tests/wheel,

Liquid   storage station sheath   bottle, waste bottle, shutdown liquid bottle 2L

cleaning   liquid bottle 0.5L

Software English/Chinese
PC Dell(CPU i5/8G DDR4/1T hard disk)+Dell   27 inches monitor
Size/   Weight 500mm*605mm*305mm   (L*W*H),35KG