●  Control Mode: Air mode and baby mode control by micro-computer

●  Temperature control range:25℃-34℃

●  Skin temperature sensor display range: 5℃-65℃

●  Correlation of displayed average value and actual average value (stable temperature condition): <0.8℃

●  Accuracy of skin temperature sensor: +0.2℃

●  Temperature uniformity of the bed: s0.8C

●  Noise level above the bed: s55dB(A)

●  Failure alarm: Over-temperature alarm; Deviation alarm, Sensor failure alarm, Fan failure alarm and Check alarm. The total irradiance for bilirubin on the effective surface area: Above case:>23umw/cm’nm (blue fluorescent lamp)

●  >30uw/cm’/nm (Blue LED)

●  Nether case:>45uw/cm/nm

●  Average value of general bilirubin radiation on the effective surface area: Above case:>21uw/cm/nm (Blue fluorescent lamp)

●  ≥23uw/cm’/nm (Blue LED)

●  Nether case: >28uw/cm’/nm

●  Uniformity of the total irradiance for bilirubin: >0.4

●  Dominant wavelength of blue light: 420nm-470nm