PA-101 Portable Ventilator

Description

APPLICATIONS

● Emergency situation

● Patient transport

 

MAIN FEATURES

● Practical functions and convenient operation.

● High-quality main components with high reliability.

● Complete alarm system.

● Instantaneous automatic pressure relief for overpressure protection.

 

ENVIRONMENTAL REQUIREMENTS

● Storage temperature: -18~50℃

● Relative humidity: ≤95% (non-condensing)

● Atmospheric pressure: 70~110kPa

● Gas source: medical oxygen at 280~ 600kPa

● Power: AC 100-240V, 50/60Hz, 30VA

● Internal power supply: Lithium-ion Battery, 12V/2200mAh

 

STANDARD CONFIGURATIONS

● Main unit

● Mask

● Active exhalation valve

● Respiratory tubes

● Expiratory control sampling tube

● Oxygen tubing

● Oxygen regulator

● Carrying bag

OPTIONAL

● Oxygen cylinder

 

MAIN PARAMETERS

● Ventilation mode: CMV

● Respiratory rate: 5 ~ 40bpm (step value 1bpm), ±10%

● I:E ratio: 1:2, ±10%

● Minute ventilation: 0 ~ >20L/min (adjustable)

● FiO2: Air mixed in (50 ~ 80%) or No Air mixed in (90 ~ 100%), ±3%

● MAWP (maximum allowable working pressure): ≤8.0kPa

● Continuous running time: 24 hours on AC power; not less than 8 hours on batteries

 

SAFETY MEASURES

● Low voltage: Medium priority alarm

● Depleted battery: High priority alarm

● High airway pressure:

● Medium priority alarm when the airway pressure is above 5.5kPa;

● Automatic pressure relief when the airway pressure exceeds 8kPa.